Cookies on this website

We use cookies to ensure that we give you the best experience on our website. If you click 'Accept all cookies' we'll assume that you are happy to receive all cookies and you won't see this message again. If you click 'Reject all non-essential cookies' only necessary cookies providing core functionality such as security, network management, and accessibility will be enabled. Click 'Find out more' for information on how to change your cookie settings.

Most pulse oximeters automatically alter the intensity of their light-emitting diodes (LEDs) according to the absorption of the finger, toe or earlobe to which they are attached. This paper investigates the effect of changing LED intensity on pulse oximeter accuracy. Our results show that the peak wavelength of a red LED typically increases by 8 nm as its intensity is increased ten-fold. To determine whether this shift introduces a significant error, a simple theoretical model based on the Beer-Lambert law is used. The model predicts that a 10:1 change in LED intensity results in a 2.5% error at 50% arterial oxygen saturation (SpO2). At high saturations (SpO2 greater than or equal to 85%) the model predicts little loss of accuracy and thus any effect due to changes in LED intensity will be apparent only at low saturations.

Type

Journal article

Journal

J Med Eng Technol

Publication Date

05/1991

Volume

15

Pages

111 - 115

Keywords

Equipment Design, Hemoglobins, Humans, Hydrogen-Ion Concentration, Models, Theoretical, Oximetry, Oxygen, Predictive Value of Tests, Reference Values