We use cookies to ensure that we give you the best experience on our website. If you click 'Accept all cookies' we'll assume that you are happy to receive all cookies and you won't see this message again. If you click 'Reject all non-essential cookies' only necessary cookies providing core functionality such as security, network management, and accessibility will be enabled. Click 'Find out more' for information on how to change your cookie settings.

A commercially available holotranscobalamin (holo-TC) radioimmunoassay (RIA) (Axis-Shield, Dundee, Scotland) was evaluated in four laboratories and compared with a holoTC ELISA run in one laboratory. The performance of the holoTC RIA assay was comparable in three of the four participating laboratories. The results from these three laboratories, involving at least 20 initial runs of "low", "medium" and "high" serum-based controls (mean holoTC concentrations 34, 60 and 110 pmol/L, respectively) yielded an intra-laboratory imprecision of 6-10%. No systematic inter-laboratory deviations were observed on runs involving 72 patient samples (holoTC concentration range 10-160 pmol/L). A fourth laboratory demonstrated higher assay imprecision for control samples and systematic deviation of results for the patient samples. Measurement of holoTC by ELISA showed an imprecision of 4-5%, and slightly higher mean values for the controls (mean holoTC concentrations 40, 70 and 114 pmol/L, respectively). Comparable results were obtained for the patient samples. The long-term intra-laboratory imprecision was 12% for the holoTC RIA and 6% for the ELISA. In conclusion, it would be prudent to check the calibration and precision prior to starting to use these holoTC assays in research or clinical practice. The results obtained using the holoTC RIA were similar to those obtained using the holoTC ELISA assay.

Original publication

DOI

10.1515/CCLM.2005.185

Type

Conference paper

Publication Date

2005

Volume

43

Pages

1058 - 1064

Keywords

Enzyme-Linked Immunosorbent Assay, Europe, Humans, Radioimmunoassay, Reproducibility of Results, Sensitivity and Specificity, Societies, Medical, Time Factors, Transcobalamins