BACKGROUND: Bioelectrical impedance analysis (BIA) and dual-energy X-ray absorptiometry (DXA) serves as common modalities for body composition assessment. This study was aimed to evaluate the agreement between BIA and DXA measures in UK Biobank. METHODS: UK Biobank participants with body fat mass (FM) and fat-free mass (FFM) estimates obtained through BIA (Tanita BC418MA) and DXA concurrently were included. Correlation between BIA and DXA-derived estimates were assessed with Lin's concordance correlation coefficients. Bland-Altman and Passing-Boblok analyses were performed to quantify the difference and agreement between BIA and DXA. Multivariable linear regression was used to identify predictors influencing the differences. Finally, prediction models were developed to calibrate BIA measures against DXA. RESULTS: The analysis included 34437 participants (female 51.4%, mean age 64.1 years at imaging assessment). BIA and DXA measurements were highly correlated (Lin's concordance correlation coefficient 0.94 for FM and 0.94 for FFM). BIA (Tanita BC418MA) underestimates FM overall by 1.84 kg (23.77 vs. 25.61, p
Journal article
Clin Nutr ESPEN
29/06/2024
63
214 - 225
Bioelectrical impedance analysis, Body composition, Dual-energy X-ray absorptiometry, Fat mass, UK Biobank