We use cookies to ensure that we give you the best experience on our website. If you click 'Accept all cookies' we'll assume that you are happy to receive all cookies and you won't see this message again. If you click 'Reject all non-essential cookies' only necessary cookies providing core functionality such as security, network management, and accessibility will be enabled. Click 'Find out more' for information on how to change your cookie settings.

In the paper a pulse oximetry model is developed using an approach which combines both theoretical and empirical modelling. The optical properties of whole blood are measured as a function of cuvette depth by transmission spectrophotometry using red (660 nm) and infra-red (950 nm) light-emitting diodes as light sources. Twersky's theoretical model gives the best fit to the experimental data. A simple theoretical model which takes into account the nonlinear relationship between optical density and cuvette depth is then used to obtain an expression for the R:IR ratio, which relates the measurement of transmission at the two wavelengths. The R:IR ratio is found to be more or less independent of cuvette depth (SD = 0.14 at 100 per cent SaO2). To validate the predictions of the theoretical model, the results of a previous experiment in which the relationship between SaO2 and the R:IR ratio was recorded using a flexible cuvette are used. The experimental values are found to lie within one standard deviation from the theoretical curve relating SaO2 and the R:IR ratio. It is argued that a reasonably accurate model for pulse oximetry which is based on whole blood and not haemoglobin solutions has been developed.

Type

Journal article

Journal

Med Biol Eng Comput

Publication Date

05/1993

Volume

31

Pages

291 - 300

Keywords

Humans, Mathematics, Models, Cardiovascular, Optics and Photonics, Oximetry, Oxygen, Pulsatile Flow